TONE
Example
TONE
Syntax
Example
TRIG
Syntax
256
Chapter 6 Command Reference
PRESET state, the multimeter uses the NRDGS command. The power-on values
for SWEEP can only be used for sub-sampling (since NRDGS does not apply
to sub-sampling).
•
You cannot use the TIMER (or SWEEP) event for AC or AC+DC voltage
measurements using the synchronous or random methods (SETACV SYNC or
RNDM) or for frequency or period measurements.
•
Query Command. The TIMER? query command returns the present time
interval, in seconds, for the NRDGS timer event.
•
Related Commands: DELAY, NRDGS, SWEEP
10 OUTPUT 722;"TRIG HOLD"
20 OUTPUT 722;"INBUF ON"
30 OUTPUT 722;"DCV 10"
40 OUTPUT 722;"NPLC .1"
50 OUTPUT 722;"AZERO OFF"
60 OUTPUT 722;"MEM FIFO"
70 OUTPUT 722;"TIMER 2"
80 OUTPUT 722;"NRDGS 10 TIMER" !10 READINGS PER SAMPLE EVENT (TIMER)
90 OUTPUT 722;"TRIG SGL"
100 END
Causes the multimeter to beep once. The multimeter then returns to the previous
BEEP mode (either OFF or ON).
TONE
Related Commands: BEEP
OUTPUT 722; "TONE"
! BEEPS
Specifies the trigger event.
TRIG [event]
event
!SUSPENDS MEASUREMENTS
!ENABLES THE INPUT BUFFER
!DC VOLTAGE, 10V RANGE
!SELECTS .1 PLC OF INTEGRATION TIME
!DISABLES AUTOZERO
!ENABLES READING MEMORY (FIFO MODE)
!SELECTS 2 SECOND INTERVAL
!TRIGGERS ONCE