TONE

Example

TONE

Syntax
Example

TRIG

Syntax
256
Chapter 6 Command Reference
PRESET state, the multimeter uses the NRDGS command. The power-on values
for SWEEP can only be used for sub-sampling (since NRDGS does not apply
to sub-sampling).
You cannot use the TIMER (or SWEEP) event for AC or AC+DC voltage
measurements using the synchronous or random methods (SETACV SYNC or
RNDM) or for frequency or period measurements.
Query Command. The TIMER? query command returns the present time
interval, in seconds, for the NRDGS timer event.
Related Commands: DELAY, NRDGS, SWEEP
10 OUTPUT 722;"TRIG HOLD"
20 OUTPUT 722;"INBUF ON"
30 OUTPUT 722;"DCV 10"
40 OUTPUT 722;"NPLC .1"
50 OUTPUT 722;"AZERO OFF"
60 OUTPUT 722;"MEM FIFO"
70 OUTPUT 722;"TIMER 2"
80 OUTPUT 722;"NRDGS 10 TIMER" !10 READINGS PER SAMPLE EVENT (TIMER)
90 OUTPUT 722;"TRIG SGL"
100 END
Causes the multimeter to beep once. The multimeter then returns to the previous
BEEP mode (either OFF or ON).
TONE
Related Commands: BEEP
OUTPUT 722; "TONE"
! BEEPS
Specifies the trigger event.
TRIG [event]
event
!SUSPENDS MEASUREMENTS
!ENABLES THE INPUT BUFFER
!DC VOLTAGE, 10V RANGE
!SELECTS .1 PLC OF INTEGRATION TIME
!DISABLES AUTOZERO
!ENABLES READING MEMORY (FIFO MODE)
!SELECTS 2 SECOND INTERVAL
!TRIGGERS ONCE