Chapter 6 Command Reference
Temperature Query. Returns the multimeter's internal temperature in degrees
Centigrade.
TEMP?
Monitoring the multimeter's tempernture is helpful to determine when to
perform autocalibration.
Related Commands: ACAL, CAL, CALSTR
10 OUTPUT 722; "TEMP?"
20 ENTER 722; A
30 PRINT A
40 END
On previous multimeters, the TERM command internally connected or
disconnected the multimeter's input terminals. The 3458 accepts the TERM
command to maintain language compatibility with these multimeters, but does
not respond since the 3458's input terminals cannot be controlled from remote.
TERM [source]
source
The source parameter choices are:
Numeric
source
Query
Parameter
Equiv.
OPEN
0
FRONT
1
REAR
2
Power-on source = none.
Default source = FRONT.
Query Command. The TERM? query command returns a response indicating
which input terminals (FRONT or REAR) are selected by the front panel
Terminals switch.
!READ TEMPERATURE
!ENTER RESULT
!PRINT RESULT
Description
Generates error message
Generates error message if Terminals switch is
set to Rear
Generates error message if Terminals switch is
set to Front