Syntax
Remarks
Example

LFREQ

Syntax
LFILTER [control]
control
The control parameter choices are:
Numeric
control
Query
parameter
Equiv.
OFF
0
ON
1
Power-on control = OFF.
Default control = ON.
Level filtering can be used when level triggering for DC voltage, direct- and
sub-sampling. The level filter can also be used to reduce sensitivity to noise for
frequency and period measurements or when making AC or AC+DC voltage
measurements using the synchronous method (SETACV SYNC command).
Query Command. The LFILTER? query command returns the present level
filter mode. Refer to "Query Commands" near the front of this chapter for more
information.
Related Commands: DCV, DSAC, DSDC, FREQ, LEVEL, NRDGS, PER,
SETACV, SYNC, SLOPE, SSAC, SSDC, SSRC, TRIG
OUTPUT 722;"LFILTER ON" !ENABLES THE LEVEL FILTER
The LFREQ command allows you to specify the A/D converter's reference
frequency or measure the line frequency and set the reference frequency to the
measured value.
LFREQ [frequency]
or
LFREQ LINE
frequency
Allows you to specify the reference frequency. The valid range for the frequency
parameter is 45 - 65 Hz, or 360 - 440 Hz. When you specify a frequency in the
range of 360 - 440Hz, the multimeter divides that value by 8. For example, if you
specify LFREQ 400, the multimeter sets the reference frequency to 400/8 = 50Hz.
Power-on reference frequency = rounded value of 50 or 60Hz (see first Remark
below).
Description
Disables the level filter; no filtering is done
Enables the level filter
Chapter 6 Command Reference

LFREQ

191