Fixedz - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

FIXEDZ

Example

FIXEDZ

Syntax
control
Remarks
Example
180
Chapter 6 Command Reference
Related Commands: NRDGS, SRQ, STB?, SWEEP, TBUFF
OUTPUT 722;"EXTOUT APER"
The FIXEDZ command enables or disables the fixed input resistance function
for DC voltage measurements. When enabled, the multimeter maintains its input
resistance at 10 megohms for all ranges. This prevents a change in input
resistance (caused by a range change) from affecting the DC voltage
measurements.
FIXEDZ [control]
The control parameter choices are:
Numeric
control
Query
Parameters
Equiv.
OFF
0
ON
1
Power-on control = OFF.
Default control = ON.
FIXEDZ remains enabled when you change from DC voltage measurements to
2-wire or 4-wire ohms measurements. Resistance measurements made with
FIXEDZ enabled will be in error because the multimeter's input resistance
represents a 10 MW resistance in parallel with the input terminals.
FIXEDZ is temporarily disabled when you change from DC voltage
measurements to AC voltage, AC+DC voltage, any type of current, frequency,
or period measurements. For example, if FIXEDZ is enabled and you change
from DC voltage measurements to AC voltage measurements, FIXEDZ
becomes disabled. When you return to DC voltage measurements, however,
FIXEDZ is once again enabled.
Query Command. The FIXEDZ? query command returns the present fixed
input resistance mode. Refer to "Query Commands" near the front of this chapter
for more information.
Related Commands: DCV, FUNC, OHM, OHMF,
OUTPUT 722; "FIXEDZ ON"
!SETS EXTOUT EVENT TO APERTURE WAVEFORM
DCV .1V,1V, 10V
Description
1000V Ranges
FIXEDZ
>10 GW
disabled
FIXEDZ
10 MW
enabled
!ENABLES FIXED IMPEDANCE
Input Resistances
DCV 100V,
Ranges
10 MW
10 MW