Slope - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

SLOPE

Remarks
Example

SLOPE

Syntax
Remarks
234
Chapter 6 Command Reference
sampling, or synchronous sampling. The parameters are:
Numeric
type
Query
Parameter
Equiv.
ANA
1
RNDM
2
SYNC
3
Power-on type = ANA.
Default type = ANA.
Bandwidth limitations vary with the conversion technique selected. See the
Specifications in Appendix A for details.
Query Command. The SETACV? query command returns the present AC
measurement method. Refer to "Query Commands" near the front of this chapter
for more information.
Related Commands: ACBAND, ACDCV, ACV, FUNC, SSRC
10 OUTPUT 722; "SETACV SYNC"
COUPLED)
20 OUTPUT 722;"ACDCV"
30 END
SLOPE is used in conjunction with the LEVEL command and specifies which
slope of the signal will be used by the level-detection circuitry.
SLOPE [slope]
slope
Selects the positive-going or negative-going slope of the input signal for use by
the level detection circuitry. The choices are:
Numeric
slope
Query
Parameter
Equiv.
NEG
0
POS
1
Power-on slope = POS.
Default slope = POS.
Query Command. The SLOPE? query command returns the present slope.
Description
Analog RMS conversion
Random sampling conversion
Synchronous sampling conversion
!SPECIFIES SYNCHRONOUS SAMPLING (DC
!SELECTS AC+DC VOLTAGE MEASUREMENTS
Description
Selects negative-going slope
Selects positive-going slope

Hide quick links: