Examples

RSTATE

Syntax
You enable a condition by specifying its decimal weight as the value parameter.
For more than one condition, specify the sum of the weights. The conditions and
their weights are:
Decimal
Bit
Weight
Number
1
0
2
1
4
2
8
3
16
4
32
5
64
6
128
7
Power-on value: If Power-On SRQ was enabled when power was removed, value
= 8; otherwise, value = 0.
Default value = 0 (no conditions enabled).
You can control the errors that will set bit 5 with the EMASK command.
The power-on SRQ bit is stored in continuous memory. All other bits are cleared
at power-on.
Query Command. The RQS? query command returns the weighted sum of all
enabled bits in the status register.
Related Commands: CSB, SPOLL (GPIB command), STB?
OUTPUT 722;"RPS 4"
OUTPUT 722;"RQS 40"
CONDITIONS
OUTPUT 722;"RQS 255"
OUTPUT 722;"RQS 0"
Recall State. Recalls a stored state from memory and configures the multimeter
to that state. States are stored using the SSTATE command.
RSTATE [name]
name
State name. A state name may contain up to 10 characters. The name can be alpha,
alphanumeric, or an integer in the range of 0 to 127. Refer to the SSTATE
command for details.
Enables Condition
Program Memory Execution Completed
Hi or Lo Limit Exceeded
SRQ Command Executed
Power-On SRQ
Ready for Instructions
Error (Consult Error Register)
Service Requested (you cannot disable this bit)
Data Available
!ENABLES THE FRONT PANEL SRQ CONDITION
!ENABLES POWER-ON SRQ (8) & !ERROR (32)
!ENABLES ALL CONDITIONS
!DISABLES ALL CONDITIONS