PRESET
218
Chapter 6 Command Reference
AZERO ON
BEEP ON
DCV AUTO
DELAY -1
DISP ON
FIXEDZ OFF
FSOURCE ACV
INBUF OFF
LOCK OFF
MATH OFF
All math registers set to 0 except:
DEGREE = 20
PERC = 1
REF = 1
RES = 50
SCALE = 1
FAST
PRESET FAST configures the multimeter for fast readings, fast transfer to
memory, and fast transfer from memory to GPIB. (Refer to "Increasing the
Reading Rate" in Chapter 4 for more information on fast measurements.)
Executing PRESET FAST executes the commands shown under PRESET NORM
with the following exceptions:
DCV 10
AZERO OFF
DISP OFF
MFORMAT DINT
OFORMAT DINT
TARM SYN
TRIG AUTO
DIG
PRESET DIG configures the multimeter for DCV digitizing (DCV digitizing is
discussed in Chapter 5). Executing PRESET DIG executes the commands shown
under PRESET NORM with the following exceptions:
DCV 10
AZERO OFF
DELAY 0
DISP OFF
TARM HOLD
TRIG LEVEL
LEVEL 0,AC
NRDGS 256,TIMER
TIMER 20E-6
APER 3E-6
MFORMAT SINT
MFORMAT SREAL
MMATH OFF
NDIG 6
NPLC 1
NRDGS 1,AUTO
OCOMP OFF
OFORMAT ASCII
TARM AUTO
TIMER 1
TRIG SYN