ABORT 7 (IFC)

ABORT 7 (IFC)

Syntax
Example

CLEAR (DCL or SDC)

Syntax
Examples

LOCAL (GTL)

Syntax
Remarks
304
Appendix B GPIB Commands
Clears the multimeter's interface circuitry.
ABORT 7
ABORT 7
!CLEARS THE MULTIMETER'S INTERFACE CIRCUITRY
Clears the multimeter, preparing it to receive a command. The CLEAR command
does the following:
Clears the output buffer.
Clears the input buffer.
Aborts subprogram execution.
Clears the status register (bits 4, 5, and 6 are not cleared if the condition(s) that
set the bit(s) still exist).
Clears the display
Disables triggering (the previous triggering mode can be resumed by sending
any multimeter command).
CLEAR 7
CLEAR 722
CLEAR 7
!CLEARS ALL DEVICES (DCL) ON THE BUS (SELECT CODE 7)
CLEAR 722
!CLEARS THE DEVICE (SDC) AT ADDRESS 22 (SELECT CODE 7)
Removes the multimeter from the remote state and enables its keyboard (provided
the keyboard has not been disabled with the multimeter's LOCK command).
LOCAL 7
LOCAL 722
If the multimeter's LOCAL key is disabled by LOCAL LOCKOUT. the LOCAL
722 command enables the keyboard, but a subsequent remote command disables
the keyboard. Sending the LOCAL 7 command, however, returns front panel
control even after a subsequent remote message.