Burst Complete - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

10 OUTPUT 722;"PRESET NORM"
20 OUTPUT 722;"MEM FIFO"
30 OUTPUT 722;"TRIG EXT"
40 OUTPUT 722;"EXTOUT RCOMP,NEG"
45
50 OUTPUT 709;"SADV EXTIN"
60 OUTPUT 709;"CHCLOSED EXT"
70 OUTPUT 709;"SCAN 201- 206"
75
80 END

Burst Complete

10 OUTPUT 722;"PRESET NORM"
20 OUTPUT 722;"MEM FIFO"
30 OUTPUT 722;"TRlG EXT"
40 OUTPUT 722;"EXTOUT BCOMP, NEG"!BURST COMPLETE EVENT, LOW-GOING TTL
50 OUTPUT 722;"NRDGS 15, AUTO"!15 READINGS PER CHANNEL
55
60 OUTPUT 709;"SADV EXTIN"
70 OUTPUT 709;"CHCLOSED EXT"
80 OUTPUT 709;"SCAN 201- 206" !SCAN CHANNELS 01 - 06 ON SCANNER IN SLOT 200
85
90 END
!DCV,NRDGS,l,AUTO, TARM AUTO, TRIG SYN
!ENABLE READING MEMORY, FIFO MODE
!TRIGGER EVENT = EXTERNAL
!READING COMPLETE EXTOUT, LOW-GOING TTL
!CONFIGURE EXTERNAL SCANNER
!ADVANCE SCANNER ON MULTIMETER'S EXTOUT SIGNAL
!OUTPUT LOW-GOING PULSE AFTER EACH CLOSURE
!SCAN CHANNELS 01- 06 ON SCANNER IN SLOT 200
!AND ADVANCE TO CHANNEL 01, STARTING THE SCAN
Figure 21. Using an external scanner
When specified, the burst complete event (BCOMP event) produces a 1 µs
pulse following completion of a group of readings. The number of readings
in a group is specified by the NRDGS or SWEEP command. The BCOMP
event can be used to synchronize an external scanner to the multimeter when
making multiple readings per scanner channel. The following program is
similar to the preceding program except that it uses the BCOMP event and
makes 15 readings on each scanner channel. Connections for this example
are shown in Figure 21.
!DCV, NRDGS 1,AUTO, TARM AUTO, TRIG SYN
!ENABLE READING MEMORY, FIFO MODE
!TRIGGER EVENT = EXTERNAL
!CONFIGURE EXTERNAL SCANNER
!ADVANCE SCANNER ON MULTIMETER'S EXTOUT SIGNAL
!OUTPUT LOW-GOING PULSE AFTER EACH CLOSURE
!AND ADVANCE TO CHANNEL 01, STARTING THE SCAN
Chapter 4 Making Measurements
113