Rmem - Agilent Technologies 3458A User Manual

Hide thumbs Also See for 3458A:
Table of Contents

Advertisement

Remarks
Example

RMEM

Syntax
Numeric
register
Query
Parameter
Equiv.
PFAILNUM
15
Power-on register = none.
Default register = DEGREE.
Math register contents are always output in the ASCII output format regardless
of the specified output format. Afterwards, the output format returns to that
previously specified (SINT, DINT, SREAL, DREAL, or ASCII).
Related Commands: MATH, MMATH, SMATH
10 OUTPUT 722;"TRIG HOLD"
20 OUTPUT 722;"MEM FIFO"
30 OUTPUT 722;"NRDGS 10"
40 OUTPUT 722;"DCV 3"
50 OUTPUT 722;"MATH STAT"
60 OUTPUT 722;"TRIG SGL"
70 OUTPUT 722;"RMATH SDEV"
80 ENTER 722;A
90 PRINT A
100 END
Recall Memory. Reads and returns the value of a reading or group of readings
stored in reading memory. RMEM leaves stored readings intact (not cleared from
memory).
RMEM [first][,count] [,record ]
first
Designates the beginning reading.
Power-on first = none.
Default first = 1.
count
Designates the number of readings to be recalled, starting with first.
Power-on count = none.
Default count = 1.
record
Register Contents
The number of reading that passed PFAIL before
a failure was encountered
!SUSPENDS TRIGGERING
!ENABLE READING MEMORY, FIFO MODE
!TEN READINGS PER TRIGGER
!DC VOLTAGE, 10V RANGE
!ENABLES STATISTICS MATH OPERATION
!TRIGGERS THE MULTIMETER ONCE
!READS STANDARD DEVIATION
!ENTERS STANDARD DEVIATION
!PRINTS STANDARD DEVIATION