4.13 Assembling the Colibri 3 LED Light Source
The present section applies to the following microscope type:
§
Skin or eye injury due to hazardous light emission
The light source belongs to Risk Group 3 as specified in IEC 62471 and emits LED radiation and
UV radiation. Skin or eye injury can result from the exposure.
4
4
4
Parts and Tools
ü
Prerequisite
Procedure
1. At the illumination mount on the rear side of the stand, remove the protective cap.
2. Loosen the clamping screw.
3. Insert the LED light source with the dovetail
Instruction Manual ZEISS Axiovert 5, Axiovert 5/7 materials | en-US | Rev. 6 | 431030-7011-101
Axiovert 5 TL FL SCB
WARNING
Avoid any eye and skin exposure to the light-emitting aperture of the light source.
Avoid exposure of skin to the radiation. Use suitable protective equipment/protective
clothing if required.
Before installing or removing the light source always make sure it is switched off.
Hex Key, 3.0 mm
The microscope is switched off.
ring into the illumination mount.
4 Installation | 4.13 Assembling the Colibri 3 LED Light Source
83