Download Print this page

Zeiss Axiovert 5 Instruction Manual page 108

Inverted microscope for research and routine microscopy
Hide thumbs Also See for Axiovert 5:

Advertisement

5 Operation | 5.7 Setting Up for Reflected Light Techniques
5.7.2 Setting Up for Reflected Light Darkfield Microscopy
Parts and Tools
ü
Prerequisite
ü
Procedure
1. Swivel the reflector module darkfield ACR P&C
2. Bring the objective position with the darkfield
3. Completely open the luminous field diaphragm.
4. Completely open the aperture diaphragm.
5. Switch off or remove neutral filters if applicable.
108
Reflector module darkfield ACR P&C for reflected light
Objective for RL darkfield
The microscope is operational.
The illumination is adjusted for reflected light brightfield microscopy [
for reflected light on the reflector turret into
the beam path.
objective into the beam path.
Instruction Manual ZEISS Axiovert 5, Axiovert 5/7 materials | en-US | Rev. 6 | 431030-7011-101
ZEISS
}
105].

Advertisement

loading
Need help?

Need help?

Do you have a question about the Axiovert 5 and is the answer not in the manual?

This manual is also suitable for:

Axiovert 5/7