Dosimag IO-Link
Endress+Hauser
13
Maintenance
13.1
Maintenance work
No special maintenance work is required.
13.1.1
Cleaning of surfaces not in contact with the medium
1. Recommendation: Use a lint-free cloth that is either dry or slightly dampened using
water.
2. Do not use any sharp objects or aggressive cleaning agents that corrode the surfaces
(displays, housing, for example) and seals.
3. Do not use high-pressure steam.
4. Observe the degree of protection of the device.
NOTICE
Cleaning agents can damage the surfaces!
Incorrect cleaning agents can damage the surfaces!
‣
Do not use cleaning agents with concentrated mineral acids, bases or organic solvents
e.g. benzyl alcohol, methylene chloride, xylene, concentrated glycerol cleaners or
acetone.
13.1.2
Cleaning of surfaces in contact with the medium
Note the following for cleaning and sterilization in place (CIP/SIP):
• Use only cleaning agents to which the materials in contact with the medium are
sufficiently resistant.
• Observe the permitted maximum medium temperature.
13.1.3
Cleaning with pigs
It is essential to take the internal diameters of the measuring tube and process connection
into account when cleaning with pigs. All the dimensions and lengths of the measuring
instrument are provided in the separate "Technical Information" document.
13.1.4
Replacing seals
The measuring instrument seals (particularly aseptic molded seals) must be replaced
periodically.
The interval between changes depends on the frequency of the cleaning cycles, the
cleaning temperature and the medium temperature.
Replacement seals (accessory) → 65
13.2
Measuring and test equipment
Endress+Hauser offers a variety of measuring and testing equipment, such as Netilion or
device tests.
Your Endress+Hauser Sales Center can provide detailed information on the services.
List of some of the measuring and testing equipment: → 52
Maintenance
47