Technical Specification - EMS COMBINATION 850 Manual

Table of Contents

Advertisement

General
Power Input
Battery
Classification (EN60601-1)
Mains Fuses
Size (height x width x depth)
Weight
Treatment Programs
Ultrasound
Frequency
Maximum Intensity
Maximum Output Power
Output Modes
Pulse Duration
Treatment Timer
Contact Monitor
Large Ultrasound Transducer
ERA
BNR
Beam Type
Small Ultrasound Transducer
ERA
BNR
Beam Type
Transducers for use with the Combination 850 are fully interchangeable and
suitable for underwater treatment (IPx7 rated).
COMBINATION 850

Technical Specification

100-240 Vac 50/60 Hz
Internal Rechargeable (NiMh)
Class 1, Type BF
2 x T630 mA (5 x 20 mm)
100 x 240 x 210 mm (including handle)
2.2 kg
16 user-defined set-ups
1.1 MHz ±5% and 3.3 MHz ±5%
2 W/cm
10 W (Mains operation only)
CW and pulsed 1:1, 1:2, 1:4 and 1:9
2 ms
0 to 20 minutes (treatment linked)
Light on transducer
5 cm
<6
Collimating
0.5 cm
<6
Collimating
13
2
2
2
ISSUE 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

90

Table of Contents