Due to the internal battery, the Vivo 3 may start if the Start/Stop button is pressed even
without the mains being connected.
Do not use the Vivo 3 with nitric oxide, helium or helium mixtures. This may affect
patient air flow and volume measurements.
If you suspect that the device has been mistreated, perform a functional check before
taking it to use. A basic functional check can be performed as described in 5.5 Performing
Start-up Checks, page 38. A complete functional check can be performed by an authorized
service technician.
NOTE
Any serious incident that has occurred in relation to this device should be reported to
the competent authority and the manufacturer.
2.2
Electricity — Warnings and Precautions
WARNING!
Risk of Electric Shock
High voltage contact may cause cardiac arrhythmia.
Doc. 007231 F-3
Do not operate the Vivo 3 if it has a damaged power cord, power supply or casing.
To avoid electrical shock, only clean the Vivo 3 according to instructions in this
manual. Do not soak or immerse the Vivo 3 into any fluids.
Use the approved power supply units only.
Use of unapproved power supply units may compromise the electrical isolation
and lead to risk of electric shock.
Do not use more than one multiple portable socket-outlet or extension cord.
If a multiple portable socket-outlet is used, it must not be placed on the floor.
The operator must not touch accessible contacts of connectors and the patient
simultaneously.
Nurse Call must only be connected to a safety extra low voltage system with an iso-
lation from AC power (Mains) voltage which complies with the requirements of
IEC 60601-1.
Safety Information 11