The Others - Breas Vivo 3 User Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

3.5.5
The Others Pages
The Others pages contain non-clinical settings and information.
3.5.5.1
Device Settings
This section describes device settings that don't affect the ventilating function of the Vivo 3.
Setting
Language
Confirm Start/Stop
Pressure Unit
Display Light
Light Intensity
3.5.5.2
Pre-use Test
The pre-use test is to verify that the machine and the patient circuit work as expected, the
default circuit setting is for a 15 mm circuit diameter. Go to Setup > Patient circuit to
change the default setting.
If Custom is selected, a pre-use test must be performed. The resistance value is saved to
configure the custom circuit.
The pre-use test is performed using a guide in the graphical user interface.
28 Product Description
User Manual
Description
Value range:
The available languages are listed with their native names
accompanied by a flag representing a country where the lan-
guage is spoken.
Default value: English (The language selection menu is dis-
played when the Vivo 3 is started for the first time.)
If set to On, the user need to confirm that the treatment is to
be stopped.
If set to Off, treatment is started and stopped directly when
pushing the start/stop button on the machine.
Default value: On
Value range:
cmH
O
2
mbar
hPa
Default value: cmH
2
Value range:
On — Always lit at the selected intensity
Auto — Always lit, with automatic adjustment of the inten-
sity with regards to the ambient light.
Delayed — The display is dimmed after about 30 seconds
(time depends on usage mode and power source). If any
button is pressed or any alarm occurs, the display is lit again.
Default value: On
Value range: 1 to 5
Default value: 5