Disposal - Breas Vivo 3 User Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

* If the Vivo 3 is used in an environment with high grades of pollen or pollutions, shorter
intervals might be required.
1
Turn off the Vivo 3 and place it on a clean dust free surface.
2 Pull out the filter holder and remove the filters.
3 If required by the interval or if visibly dirty, wash
the grey coarse filter:
1. Wash the filter using warm water and a mild
2. Rinse thoroughly.
3. Dry the filter by first squeezing it in a towel
4. Check that the filter is undamaged and completely dry before reinstalling it.
4 When reinstalling the air inlet filters in the filter
holder: first install the grey coarse filter, then the white
fine filter.
5 Reinstall the filter holder on the Vivo 3.
8.3

Disposal

The ventilator, any accessories and all replaced parts must be disposed of and recycled in
accordance with the local environmental regulations regarding the disposal of used equip-
ment and waste. Contact your service provider for information regarding the disposal
procedure.
NOTE
Batteries used with the ventilator shall be recycled in accordance with the local environ-
mental regulations.
Doc. 007231 F-3
soap.
and then letting it dry in the air. Do not wring
the filter.
Cleaning and Maintenance 71