To Select Harmonic Amplitude

After entering the harmonic setting interface, tap the Harm Ampl parameter input
field to set the amplitude of each order of harmonic.
1) No.: selects the sequence number of the harmonic to be set.
2) Harm Ampl: sets the amplitude of the specified order of harmonic. Use the
numeric keypad to input the amplitude value, and then select the desired unit.
Tap "Ok". The amplitude units available are Vpp, mVpp, Vrms, mVrms, and dBm
(invalid in HighZ).

To Set Harmonic Phase

After entering the harmonic setting interface, tap the Harm Phase parameter input
field to set the phase of each order of harmonic.
1. No.: selects the sequence number of the harmonic to be set.
2. Harm Phase: sets the phase of the specified order of harmonic. Use the
numeric keypad to input the phase value, and then select the desired unit "°".
Tap "Ok".

Example: To Output Harmonic

This section introduces how to output 2th order of harmonic (harmonic amplitude: 2
Vpp, harmonic phase: 30°) and 4th orders of harmonic (harmonic amplitude: 1 Vpp,
harmonic phase: 50°) from the [CH1] connector.
1. Select the output channel: Press Output1 on the front panel or tap the
channel output configuration status bar
the channel is indicated in red in the status bar.
DG2000 User Guide
to select CH1. At this time,
RIGOL
2-25