Trigger[:Sequence]:Fmt:mask:edit; Trigger[:Sequence]:Fmt:mask:relative:amplitude - Rigol SA5000 Series Programming Manual

Vector signal analysis
Table of Contents

Advertisement

Chapter 2 Command System
Queries the mask to be used for the current trigger.
Parameter
Name
Type
——
Keyword
Remarks
This command is only valid when you select FMT to be the trigger source.
Return Format
The query returns UPP, LOW, or BOTH.
Example
The following command sets the mask to be used for the current trigger to Upper.
:TRIGger:SEQuence:FMT:MASK UPPer
The following query returns UPP.
:TRIGger:SEQuence:FMT:MASK?

:TRIGger[:SEQuence]:FMT:MASK:EDIT

Syntax
:TRIGger[:SEQuence]:FMT:MASK:EDIT UPPer|LOWer
:TRIGger[:SEQuence]:FMT:MASK:EDIT?
Description
Sets the mask type that is currently viewed/edited.
Queries the mask type that is currently viewed/edited.
Parameter
Name
Type
——
Keyword
Remarks
This command is only valid when you select FMT to be the trigger source.
Return Format
The query returns UPP or LOW.
Example
The following command sets the mask that is currently viewed/edited to Upper.
:TRIGger:SEQuence:FMT:MASK:EDIT UPPer
The following query returns UPP.
:TRIGger:SEQuence:FMT:MASK:EDIT?

:TRIGger[:SEQuence]:FMT:MASK:RELative:AMPLitude

Syntax
:TRIGger[:SEQuence]:FMT:MASK:RELative:AMPLitude OFF|ON|0|1
:TRIGger[:SEQuence]:FMT:MASK:RELative:AMPLitude?
Description
Sets whether the amplitudes of the mask points are coupled to the reference level of the instrument.
VSA Programming Guide
Range
UPPer|LOWer|BOTH
Range
UPPer|LOWer
Default
UPPer
Default
UPPer