or use the numeric keypad to set the DC bias voltage.
Probe Information
Press About to view information about the probe currently connected, such as
the manufacturer, model, serial number, and the last calibration date.
Probe Calibration
Press Calibration to start zero calibration automatically for the probe.

Amplitude Unit

Select the amplitude display unit for the current channel. The available units are W, A,
V, and U. When the unit is changed, the unit related to the relevant functions of the
channel will change accordingly.
Press 1 → More → Unit, press Unit continuously or rotate the multifunction knob
to select a unit. You can also enable the touch screen to select the desired unit
with touch gestures. The default unit for the value is V.

Channel Delay

When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring relatively greater errors (zero offset). MSO8000 allows you to
set a delay time for calibrating the zero offset of the corresponding channel. Zero
offset is defined as the offset of the crossing point of the waveform and trigger level
line relative to the trigger position, as shown in Figure 2-2.
Press 1 → More → Ch-Ch Skew, rotate the multifunction knob
2-10
Figure 2-2 Zero Offset
Chapter 2 To Set the Vertical System
or use the
MSO8000 User Guide