Using the Pattern Generator
Set Up the Timing Analyzer Trigger Term
Set Up the Timing Analyzer Trigger Term
The trigger term 'a' is set to 1 so that the analyzer will trigger on the "walking
ones" provided by the pattern generator.
Press the
1
Highlight the field to the right of term "a", and type a 1 into the field.
2
Press the
3
Highlight the
4
Select
Highlight the field to the right of >, and turn the knob to select 8 ns
5
for the pattern duration. Press the
The analyzer is set up to
trigger on the
occurrence of a 1.
The a term holds the
value 1.
Setting Up the Timing Analyzer Trigger Term
Result
The logic analyzer is set up to trigger on the occurrence of a 1.
9-8
Trigger
MENU key.
key.
Done
1
field in the Timing Sequence Levels box, and press the
key.
Get other manuals https://www.bkmanuals.com
Done
key.