Error Indications - Siemens SIPROTEC 5 V7.80 Operation

Table of Contents

Advertisement

I/O modules
A list of the input/output modules being used can be found under the menu item I/O Modules. You can
select each individual printed circuit board assembly and, for example, have the slot or the terminal posi-
tion displayed.
Transformer modules
If using transformer modules, this menu item displays all available information about these modules (for
example, slot, board type or serial number)
Fatal Error, the Device Goes into the Fallback Mode
Certain fatal device errors lead to the device falling immediately into the fallback mode. Fatal device errors are
errors that cannot be resolved by a restart of the device. An indication displays the type of error. From this,
you can derive additional steps (for example, contacting the repair department). The device goes permanently
out of operation, a failure is avoided. In fallback mode, minimal operation of the device via the on-site opera-
tion panel and DIGSI 5 is possible. In this way you can still read information from the diagnostic log, for
example.
Life contact
Red error LED
Group-Warning Indication
Pickup of the following supervisions with entry of the device into the fallback mode does not allow output of
normal supervision indications. The entry of the device into the fallback mode thus also does not lead to the
activation of the group warning indication.
10.2.4

Error Indications

If the SIPROTEC 5 device is outside a normal operating mode (for example, device is in commissioning mode or
in simulation mode), this is shown by an indication. By default this indication is prerouted to LED 16 and
makes the red LED flash.
If you have exited one of the modes incorrectly (for example, by pulling out the DIGSI 5 PC), the red LED
flashes and after 4 hours the
This function is realized in the SIPROTEC 5 device via a predefined CFC chart. This means that you can change
this behavior in case of doubt. To do this, change or delete the CFC chart.
Overview of Errors
For each error indication on the on-site operation panel, a specific error code is issued additionally (see
Figure 10-3
(2)). This error code is helpful for further analysis by the repair service.
Device-Diagnostic Log
Memory failure (recoverable): Reset initiated.
PCB link failure (non recoverable):
Check the module configuration and interconnection.
If the failure remains, contact the repair service.
Binary-output failure at module 1-12:
Contact the repair service.
Hardware failure at module 1-12:
Contact the repair service.
Wrong module 1-12 detected:
Synchronize the hardware configuration of the device with DIGSI.
Wrong plug-in module detected at position E/F/M/N/P:
Synchronize the hardware configuration of the device with DIGSI.
SIPROTEC 5, Operation, Manual
C53000-G5040-C003-9, Edition 06.2018
is terminated in fallback mode
is activated in fallback mode
Device Ready indication will go.
Maintenance, On-the-Spot Assistance and Test
10.2 Error Search and Correction
219