Power On/Off; Scanner Power On - GE LOGIQ P9 Service Manual

Hide thumbs Also See for LOGIQ P9:
Table of Contents

Advertisement

GEDRAFT
D
5604324, R
IRECTION
EVISION
4-3-3

Power On/Off

After turning off a system, wait at least 10 seconds before turning it on again.
NOTICE
The system may not be able to boot if power is recycled too quickly.
!! NOTICE:
4-3-3-1

Scanner Power On

1) Connect the Main Power Cable to the back of the system.
2) If not already done, screw on the pull-out protection of the mains power cable with the 2 screws.
3) Connect the Main Power Cable to a hospital grade power outlet with the proper rated voltage. Never
use an adapter that would defeat the safety ground.
4) Switch ON the Circuit Breaker at the rear of the system. (see:
Figure 4-11 Circuit Breaker, pull-out protection and main power cable at rear of system.
When AC power is applied to the scanner, the On/Off button on the control panel illuminates BLUE
NOTICE
indicating that the system (including back-end processor) is in Standby Mode.
!! NOTICE:
5) Hold down the On/Off button (see:
NOTE:
The power for on-board peripheral auxiliary equipments are commonly switched with the ON/OFF
button. The power switch of any attached printer(s) needs to be in ON position before starting the
system. However, be aware some auxiliary equipment may switch itself to standby mode (e.g., Color
video printer) and must therefore be switched on separately.
11
DRAFT (J
Mains Power Cable
Chapter 4 - Functional Checks
24, 2019)
ANUARY
Pull Out Protection
(Metal Bracket)
Figure
4-12) on the control panel for ~3 seconds.
LOGIQ P9/P7
S
ERVICE
Figure
4-10)
Circuit Breaker
M
ANUAL
4-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Logiq p7

Table of Contents