GEDRAFT
D
5604324, R
IRECTION
EVISION
Section 8-30
Replacement of the Probe Holder
8-30-1
Manpower
1 Person, 5 minutes.
8-30-2
Tools
No tool is required.
8-30-3
Removal Procedure
1) Simply pull out Probe Holder from its location.
NOTE:
This procedure applies to all locations of probe holder.
8-30-4
Installation Procedure
Parts to be installed in reverse order of removal.
8-30-5
Functional Check
Visual inspection Only.
11
DRAFT (J
Figure 8-180 Removal of Probe Holder
Probe holder
Figure 8-181 Probe Holders
Chapter 8 - Replacement Procedures
24, 2019)
ANUARY
4D Probe holder
LOGIQ P9/P7
S
M
ERVICE
ANUAL
8-153