4) Run the OPIO Test tool (ScFpApiTest.exe) in the desktop screen and then click the "Op-panel
Open" button of test tool. (Program path: C:\scanner\target\bin\ScFpApiTest.exe)
5) Check on the "Keyboard Test on" check box.
6) Check that pressed key is displayed except for 'Fn' key, when AN Keyboard is pressed.
7) End ScFpApiTest by click the "Op-Panel Close" button.
8-16-7
Functional Check
System Exterior Visual Check
Power On/Off
System Integration Checks
OPIO Interface Check
8-100
Section 8-16 - Replacement of Key Caps (by special native language keys)
11
DRAFT (J
Figure 8-104
Figure 8-105
Test
section 4-3-1
section 4-3-3
section 4-3-5
section 4-3-8
24, 2019)
ANUARY
Refer to
Control panel visual check only
section 4-3-5-1 OPIO Test only