Understanding Option 012 (Lo In/Out For Phase Coherency) With Multiple Baseband Generator Synchronization; Configuring The Option 012 (Lo In/Out For Phase Coherency) With Mimo - Agilent Technologies X Series User Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Basic Digital Operation (Option 653/655/656/657)

Understanding Option 012 (LO In/Out for Phase Coherency) with Multiple Baseband Generator Synchronization

Understanding Option 012 (LO In/Out for Phase Coherency) with Multiple
Baseband Generator Synchronization
NOTE
This section assumes that the previous section on Multiple Baseband Generator
Synchronization has been read and understood. If not, refer to
Synchronization" on page 225
The MXG/EXG with Option 012, enables 2x2, 3x3, or 4x4 MIMO configurations to share a common
external LO signal to create a phase coherent system (refer also, to
Synchronization" on page
RF phase coherency may not be needed for general STC/MIMO receiver testing, since a MIMO receiver
perceives any phase differences between the sources as part of the channel conditions and correct for
them. But, RF phase coherency might be desirable for certain applications such as R&D on
beamforming systems.

Configuring the Option 012 (LO In/Out for Phase Coherency) with MIMO

The Agilent BNC cable, part number 10502A, is the recommended cable for the standard multi–BBG
synchronization setup, and is recommended for Option 012 too (see also
Additionally, for the 2x2, 3x3, and 4x4 MIMO connections from the LO IN and LO OUT to the splitter,
additional cables are required (refer to
page
235).
NOTE
Agilent recommends the LO Output be covered when not in use.
When the LO In/Out jumper cable is removed and the instrument is in Dual ARB mode, the
instrument is unleveled and the instrument displays an Unlevel error message.
All test equipment requires a 12 hour warm–up period to ensure accurate performance.
The phase coherent configuration requires the following:
• The recommended LO input drive level should be in the 0 to 6 dBm range.
NOTE
The 0 to 6 dBm LO input drive level ensures the instruments will operate over the full
frequency and over the full 0 to 55 ambient temperature range
• The I/Q calibration and the self–test must be performed with the LO In/Out jumper cable in
place. Where the I/Q calibration cannot be run, the baseband offset can be manually adjusted to
minimize the I/Q offsets.
• The phase coherency feature only applies to the Dual ARB modulation mode.
• All cables from the splitter output to the instrument inputs should be of equal lengths.
1
LO input power requirements vary with temperature; power <0 dBm may work at 20–30 degree ambient
temperature conditions. Refer to the Data Sheet.
232
before continuing.
225).
Table 8-
"Multiple Baseband Generator
5,
Figure 8- 30 on page
Agilent X-Series Signal Generators User's Guide
"Multiple Baseband Generator
Figure 8- 29 on page
234, and
Figure 8- 31 on
1
.
229).

Advertisement

Table of Contents
loading

Table of Contents