Configuring The Lf Output With An Internal Modulation Source - Agilent Technologies X Series User Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Using Analog Modulation (Option UNT)
Configuring the LF Output (Option 303)
The RF On/Off hardkey does not apply to the LF OUTPUT connector.

Configuring the LF Output with an Internal Modulation Source

In this example, the internal FM modulation is the LF output source. See
NOTE
Internal modulation (Int Monitor) is the default LF output source.
Configuring the Internal Modulation as the LF Output Source
1. Press Preset.
2. Press the FM/ΦM hardkey.
3. Press FM Dev > 75 > kHz.
4. Press More > Setup FM Source > FM Rate > 10 > kHz.
5. Press Return > Return > FM Off On.
You have set up the FM signal with a rate of 10 kHz and 75 kHz of deviation. The FM annunciator is
activated indicating that you have enabled frequency modulation.
Configuring the Low Frequency Output
1. Press the LF Out hardkey.
2. Press LF Out Amplitude > 3 > V.
3. Press LF Out Off On.
You have configured the LF output signal for a 3 volt sine wave (default wave form) output which is
frequency modulated using the Int Monitor source selection (default source).
Figure 4-3 Configure the LF Out Source with FM
FM and LF annuciators indicate Frequency Modulation is the LF Out source
82
LF Out using the Int Monitor source (default selection).
Figure 4-
FM Modulation as the LF Out Source
For details on each key, use key help
as described on
Agilent X-Series Signal Generators User's Guide
3.
page 44
.