Modifying The Coefficients - Agilent Technologies X Series User Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Basic Digital Operation (Option 653/655/656/657)
Modifying a FIR Filter Using the FIR Table Editor
Figure 8-24
For details on each key, use key
help as described on
page
44.
7. Press Return.

Modifying the Coefficients

1. Using the front panel arrow keys, highlight coefficient 15.
2. Press 0 > Enter.
3. Press Display Impulse Response.
Figure 8-25
For details on each key, use key
help as described on
page
44.
Refer to
Figure 8-
it indicates that a coefficient value is missing, resulting in an improper Gaussian response).
4. Press Return.
5. Highlight coefficient 15.
6. Press 1 > Enter.
222
25. The graphic display can provide a useful troubleshooting tool (in this case,
Agilent X-Series Signal Generators User's Guide