Configuring the LF Output (Option 303)

The signal generator has a low frequency (LF) output. The LF output's source can be switched
between an internal modulation source or an internal function generator.
Using internal modulation (Int Monitor) as the LF output source, the LF output provides a replica of
the signal from the internal source that is being used to modulate the RF output. The specific
modulation parameters for this signal are configured through the AM, FM, or ΦM menus. The internal
source (AM, FM, or ΦM) must be configured for the LF Out to provide a signal.
Using function generator as the LF output source, the function generator section of the internal
modulation source drives the LF output directly. Frequency and waveform are configured from the LF
output menu, not through the AM, FM, or ΦM menus. You can select the waveform shape from the
following choices:
Available internal waveforms include:
Sine
sine wave with adjustable amplitude and frequency
Triangle
triangle wave with adjustable amplitude and frequency
Square
square wave with adjustable amplitude and frequency
Pos Ramp
positive going ramp with adjustable amplitude and frequency
Neg Ramp
negative going ramp with adjustable amplitude and frequency
Pulse
pulse with adjustable period and width

LF Out Modulation Sources

The signal generator provides the following modulation input sources. Internal modulation sources
generate the five waveforms listed above unless noted otherwise.
uses AM, FM or ΦM settings.
Int Monitor
Func Gen 1
waveforms from the internal function generator
Func Gen 2
waveforms from the internal function generator
Dual Func Gen
dual waveforms with individually adjustable frequencies and a percent- of-
peak- amplitude setting for the second tone. Available on instruments with Option
303.
Swept Func Gen
swept waveforms with adjustable start and stop frequencies, sweep time, and
sweep trigger settings. Available on instruments with Option 303.
Noise Gen 1 & 2
noise with adjustable amplitude generated as a peak- to- peak value (RMS value is
approximately 80% of the displayed value). Uniform and Gaussian distribution is
available. Available on instruements with Option 303.
DC
selects a DC voltage level as the LF output BNC source
NOTE
The LF Out Off On softkey controls the operating state of the LF output. However when the LF
output source selection is Int Monitor, you have three ways of controlling the output. You can
use the modulation source (AM, FM, or ΦM) on/off key, the LF output on/off key, or the Mod
On/Off softkey.
Agilent X-Series Signal Generators User's Guide
Using Analog Modulation (Option UNT)

Configuring the LF Output (Option 303)

81