Trigger Source - Agilent Technologies X Series User Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Basic Digital Operation (Option 653/655/656/657)
Triggering a Waveform
• Segment Advance mode plays a segment in a sequence only if triggered. The trigger source controls
segment–to–segment playing (see
received during the last segment loops play to the first segment in the sequence.
• Gated mode triggers the waveform at the first active triggering state, then repeatedly starts and
stops playing the waveform in response to an externally applied gating signal. See
Triggering
on page 182.

Trigger Source

Mode >
Dual ARB >
Trigger Source
Neg = the signal generator
responds during the trigger signal
low state.
Pos = the signal generator
responds during the trigger signal
high state.
Key inactive until Ext
Delay is enabled On.
For details on each key, use key help
as described on
page
44.
External Trigger Polarity
• In Continuous, Single, and Segment Advance modes, use the Ext Polarity softkey to set the external
trigger polarity.
• In Gated mode, the Active Low and Active High softkeys
trigger polarity.
180
Example: Segment Advance Triggering
on page 181). A trigger
Patt Trig In BNC connector
(see
page
17)
Aux I/O connector
(see
page
18)
(page
179) determine the external
Agilent X-Series Signal Generators User's Guide
Example: Gated