Setting A Frequency Multiplier - Agilent Technologies X Series User Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Optimizing Performance
Using an Output Offset, Reference, or Multiplier
Parameter
Output Frequency:
To set a new frequency or amplitude reference, turn the frequency reference off, and then follow the
steps above.

Setting a Frequency Multiplier

Using a frequency multiplier, the signal generator can display a frequency that is the multiple
(positive or negative) of the output value.
Displayed Value = multiplier value × output frequency
Output Frequency = displayed value ÷ multiplier value
To set a frequency multiplier:
1. Press Frequency > Freq Multiplier > multiplier value > x.
2. Set the desired frequency.
The display equals the output frequency times the multiplier value.
Indicates that a frequency multiplier is on
Examples
Parameter
Frequency Multiplier:
Entered (and displayed) Value:
124
Example
Example
Example
#1
#2
#3
52 MHz
48 MHz
1 GHz
Example
Example
Example
#1
#2
#3
−3
3
4
−600 MHz
600 MHz
8 GHz
Comments
The signal generator alerts you if the output frequency or
amplitude is out of range.
Comments
The multiplier range can be set from:
+0.001 to +1000
–1000 to –0.001
Agilent X-Series Signal Generators User's Guide