Setting Waveform Runtime Scaling - Agilent Technologies X Series User Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Setting Waveform Runtime Scaling

Runtime scaling scales the waveform data during playback; it does not affect the stored data. You can
apply runtime scaling to either a segment or sequence, and set the scaling value either while the ARB
is on or off. This type of scaling is well suited for eliminating DAC over–range errors. Runtime scaling
adjustments are not cumulative; the scaling value is applied to the original amplitude of the
waveform file. There are two ways to save the runtime scaling setting: by using the save function
(page
69) and by saving the setting to the file header
value with the waveform file, saving with the Save function stores the value as the current
instrument setting.
Use this example to learn how to scale the currently selected waveform.
1. Select the waveform to which you want to apply scaling:
a. Press Mode > Dual ARB > Select Waveform.
b. Highlight the desired waveform (segment or sequence).
c. Press Select Waveform.
2. Play the selected waveform: Press ARB Off On until On highlights.
3. Set the Waveform Runtime Scaling value:
a. Press ARB Setup > Waveform Runtime Scaling.
b. Enter a scaling value.
The signal generator automatically applies the new scaling value to the waveform. There is no
single value that is optimal for all waveforms. To achieve the maximum dynamic range, use
the largest scaling value that does not result in a DAC over–range error.
c. Press Return.
Agilent X-Series Signal Generators User's Guide
Basic Digital Operation (Option 653/655/656/657)
(page
157). Saving to the file header saves the
Scaling a Waveform
197

Advertisement

Table of Contents
loading

Table of Contents