Recalling and Applying a User Flatness Correction Array

The following example assumes that a user flatness correction array has been created and stored. If
not, perform the
Example: A 500 MHz to 1 GHz Flatness Correction Array with 10 Correction
Values
on page 102.
1. Preset the signal generator.
2. Recall the desired User Flatness Correction file:
a. Press AMPTD > More > User Flatness > Configure Cal Array > More > Preset List > Confirm Preset.
b. Press Load/Store.
c. Highlight the desired file.
d. Populate the user flatness correction array with the data contained in the selected file:
Press Load From Selected File > Confirm Load From File.
The user flatness correction array title displays User Flatness: Name of File.
3. Apply the correction data in the array to the RF output: Press Return > Flatness Off On to On.
Agilent X-Series Signal Generators User's Guide
Optimizing Performance
Using Flatness Correction