Basic Operation
Configuring a Swept Output

Routing Signals

Sweep > More > More > Route Connectors

Step Sweep

Step sweep provides a linear or logarithmic progression from one selected frequency, or amplitude, or
both, to another, pausing at linearly or logarithmically spaced points (steps) along the sweep. The
sweep can progress forward, backward, or be changed manually.
Figure 3-5 Signal Routing Softkeys
These softkeys are
only available for
the vector models
with a BBG.
Opens a menu for
routing Markers 1
through 4 to the
rear panel
baseband BNC
connectors.
52
Routes Step or
List Sweep
signals.
Routes non Step
or List Sweep
signals (i.e. SCPI
command
development.
Source Settled is
not affected by
Trig Out Polarity
Neg Pos softkey
selection.
Routes a signal to the Trig Out (Trig 1 or Trig 2) rear
panel connectors as follows:
None: No signals are routed to Trig 1 or Trig 2 BNCs
Sweep Trigger Out: Sweep Trigger Out signal (trigger
handshake)
Source Settled: Source Settled signal
Pulse Sync: Pulse Modulation Synchronization signal
Pulse Video: Pulse Modulation Video signal
LXI: LXI Trigger Output
Sweep Run: Sweep Run signal (sweep is running)
Other Trigger BNC: The signal on the other Trigger
BNC
Pulse BNC: The signal on the PULSE IN BNC
Swept Func Done: Swept Function Sweep Done signal
For details on each key, use key help as described on
Agilent X-Series Signal Generators User's Guide
page
44.