Removing the breathing circuit

1 Disconnect the breathing hoses from the
breathing system.
2 Disconnect the various parts of the hose system
(breathing hoses, Y-piece, connector, and op-
tional Y-piece filter). The filter on the Y-piece is
not reusable and can be disposed of with ordi-
nary domestic waste.
Note the regulations of the hospital for infec-
tious patients!
Note the instructions for use.
Instructions for use Primus SW 4.5n
Cleaning, Disinfection and Sterilization
3 Prepare the parts for conditioning in a cleaning
and disinfecting machine.
CAUTION
Risk of component damage
If mishandled, the spiral ribbing on the breathing
hoses can become detached from the sleeve.
Breathing hoses with damaged spiral ribbing can
easily be kinked and interrupt the flow of gas!
When attaching or removing the breathing hoses,
always hold them by the connection sleeve and
not by the spiral ribbing! Always check the breath-
ing hoses for damage prior to use. Damaged
breathing hoses must be replaced.

Removing the microbial filter (optional)

On the sleeve of the microbial filter:
1 Pull the filter off the nozzle.
2 Prepare the microbial filter for conditioning ac-
cording to the corresponding instructions for
use.