Setting The Alarm Tone; Alarms In The Standby Mode - Dräger Primus SW 4.5n Instructions For Use Manual

Anesthesia workstation
Table of Contents

Advertisement

Volume Mode,
Volume AF,
Pressure Mode,
Pressure Support
P
PEAK +5
AW
upper alarm limit
or
[hPa (cmH
O)]
PLAT +10,
2
whichever is greater.
P
0.6 x (PLAT – PEEP)
AW
+ PEEP – 1,
lower alarm limit
but at least 3
[hPa (cmH
O)]
2
To restore individual alarm limits for MV and P
see "Setting alarm limits" on page 167.
To restore all default alarm limits:
see "Operation" on page 90.

Setting the alarm tone

The volume of the alarm tone can be set in the con-
figuration menu, see page 163.

Alarms in the Standby mode

All technical alarms, e.g., failure of device compo-
nents and a number of special operating states, are
also indicated by Primus when in Standby.
A corresponding message appears in the alarms
field, but without any acoustic warning.
Instructions for use Primus SW 4.5n
:
AW
Alarms
135