Establishing A Connection To Fpm - Primes FocusParameterMonitor FPM 60 Original Instructions Manual

Table of Contents

Advertisement

9.3.3

Establishing a connection to FPM

1. Start the LDS.
2. Open the Communication > Free Communication dialog window.
3. Select "TCP" in the Mode field (the "Second IP" option must not be activated!).
4. In the TCP field, enter the IP address.
5. Click the Connect button ("Connected" appears in the bus monitor).
6. Check the Write bus log box (the log can be very useful when encountering problems):
The log is stored in the installation directory of the LDS.
The file name is buspro.log.YYYY.MM.DD (YYYY.MM.DD = date the file was created).
7. Click the Save button (the configuration is saved and does not need to be re-entered when restarting the LDS).
8. Click on the Search PRIMES Devices button:
If a device is found, the dialog windows for the measurement settings are opened.
TCP
IP address
Fig. 9.3:
Establish a connection to the PC in the Free Communication dialog window
The FPM has a fixed IP address that is specified on the type plate:
If the FPM is connected directly to the PC, enter the fixed IP address in the menu Communication
> Free Communication.
If the FPM is connected via a network, the FPM retrieves a variable IP address in the network
for around one minute.
This variable IP address can be read off with the provided software, "PrimesFindlp" and entered
into the Communication > Free Communication.
To connect the FPM to the network using the fixed IP address, first turn on the FPM and then
connect the network cable to the FPM.
Then enter the fixed IP address in the menu Communication > Free Communication.
40
192 . 168 . 116 . 119
-- Connecting to Device ip 192.168.116.80 port 6001.
--CONNECTED to 192.168.116.119.6001
FocusParameterMonitor FPM
Bus protocol
Revision 06 EN 02/2023