Measuring
12.2.4

Timing diagram caustic measurement

Fig. 12.4:
Timing diagram caustic measurement
t12:
Before the start of the first single plane measurement, "Caustic Measurement" must be set and then
only deactivated once "Caustic Measurement Finished" has been set. If several caustics are to be
measured in succession, the flag can also remain permanently set.
t13:
The caustic variables (plane index and Z position) must be set just like the single plane variables
before the "Start Measurement" flag is set. This applies to every single measurement plane.
The parameters should be deleted only once "Start Acknowledge" has been deactivated again, and
the Environmental variable should even be retained until "Measurement Finished" is set. As soon as
the caustic variables have been imported, "Caustic Measurement Running" is set.
t14:
When the last single plane has been measured and the corresponding calculations have been
performed („Measurement finished" set), „Calculate Caustic" can be set. "Caustic Measurement
Running" will then be deactivated.