t15:
Once the caustic fit and calculation of the beam parameters have been performed, "Caustic
Measurement Finished" is set. Then the results ("Caustic Variables") will also be available.
If errors occur, "Caustic Measurement Failure" is set, the cause of the failure is coded in the logs
identifier for caustic measurement errors/warnings.
t16:
"Calculate Caustic" must be deactivated before the next caustic measurement (meaning "Caustic
Measurement" is set) can be initiated.
When completing a caustic measurement, "Calculate Caustic" must be set, even in case of error,
and then deactivated again so that a new caustic measurement can be initiated.
12.2.5

Timing diagram combined measurement

Triggered focus and power measurement.
Fig. 12.5:
Timing diagramm combined measurement
72
FocusParameterMonitor FPM
Revision 06 EN 02/2023