Measuring Procedure Caustic Measurement - Primes FocusParameterMonitor FPM 60 Original Instructions Manual

Table of Contents

Advertisement

12.1.4

Measuring procedure caustic measurement

System
Sets command.caustic_measurement
Sets parameter for caustic and for
measurement
(256.0; 256.3; 292; 296;
Sets command.start (256.4)
"If status.acknowledge == 1:
deletes command.start" (256.4)
Repeat from "Sets parameter for caustic and
for measurement", for all planes
Sets command.calculate_caustic (256.5)
Reads out focus results/ error identifiers
Deletes command.calculate_caustic (256.5)
NOTICE! Setting and deleting command.
calculate_caustic must also be performed if
the caustic measurement is to be aborted (for
example, because plane results are invalid).
64
FPM
(257.6)
298)
Deletes status.idle (282.4)
Deletes status.measurement_finished (282.2)
Deletes errorflags and -identification
Reads parameters for the measurement (from the setup or the profibus)
Sets status.caustic_measurement_running (283.0)
Sets status.ready_for_measurement (282.0)
Sets status.acknowledge (282.6)
Deletes status.acknowledge (282.6)
Sets status.measurement_running (282.1)
Starts the measurement
When the measurement is finished: deletes status.measurement_running
(282.1)
Deletes status.ready_for_measurement (282.0)
Calculates parameters
Enters parameters into Profibus register or error identifiers
Sets status.measurement_finished (282.2)
Sets status.idle (282.4)
Deletes status.caustic_measurement_running (283.0)
Carries out caustic fit, calculates beam parameter
Enters parameters into Profibus register or error identifiers
Sets status.caustic_measurement_finished (283.1)