Timing Diagram Focus Measurement (Triggered) - Primes FocusParameterMonitor FPM 60 Original Instructions Manual

Table of Contents

Advertisement

Measuring
12.2.3

Timing diagram focus measurement (triggered)

The following timing diagram shows a triggered measurement. This is the standard case.
Fig. 12.3:
Timing diagram focus measurement (triggered)
t0:
The measuring parameters (setup no. or detailed measuring parameters) must be set before the
"Start" flag is set.
t1:
To confirm that the start command and measuring parameters have been received, the "Start Ack-
nowledge" flag is set and "Idle" and "Finished" deactivated. Likewise, the results log, the failure flags,
and the failure identifiers are also deactivated.
t2:
Then the "Start" flag can be deactivated.
t3:
Once "Start" has been deactivated again, "Start Acknowledge" is deactivated and "Measurement
Running" set.
t4:
Once "Start Acknowledge" has been deactivated again, this means that the measuring parame-
ters have been imported; they can then be modified as desired (e.g. in preparation for the next
measurement).