Maintenance and Service

14
Maintenance and Service
14.1

Maintenance intervals

The operator is responsible for determining the maintenance intervals of the measuring device.
PRIMES recommends a maintenance interval of 12 months for inspection, validation and calibration.
If the device is used sporadically (less than once a day), the maintenance interval can be extended up to 24
months.
Please note that the safety and warning functions of the device must be checked regularly.
14.2

Cleaning

Cleaning the device surface
1. After a measurement let the device cool down for an adequate period of time.
2. First clean the device surface with clean and oil-free compressed air.
3. Close all device openings.
4. For further cleaning, use a mixture of distilled water and isopropanol in a ratio of approx. 5:1.
Use lint-free cleaning cloths that do not cause scratches.
5. If these steps are not sufficient, please contact PRIMES or your PRIMES distributor.
14.3

Spare parts

The following spare parts are available for the FPM:
Article
Protective window cartridge
Protective window
(Diameter 15 mm; thickness. 1,5 mm)
Maintenance hatch with Ethernet port
Revision 06 EN 02/2023
Order Number
825-010-002
825-010-006
825-010-007