12.1.2

Measuring procedure focus measurement (untriggered)

System
Sets parameters for measurement
256.3) or (A257.0 - 258.0 / 260 - 299)
Sets command.start (256.4)
If status.ready_for_measurement ==1: turns
on the laser
If laser is on and status.acknowledge == 1:
deletes command.start" (256.4)
Turns off the laser (after the necessary irradia-
tion time)
Reads out results/error identifiers
62
FPM
(256.0 -
Deletes status.idle
Deletes status.measurement_finished (282.2)
Deletes errorflags and -identification
Reads parameters for the measurement (from the setup or the profibus)
Sets status.ready_for_measurement (282.0)
Sets status.acknowledge (282.6)
Deletes status.acknowledge (282.6)
Sets status.measurement_running (282.1)
Starts the measurement
When the measurement is finished: deletes status.measurement_running
(282.1)
Deletes status.ready_for_measurement (282.0)
Calculates parameters
Enters parameters into Profibus register or error identifiers
Sets status.measurement_finished (282.2)
Sets status.idle (282.4)
FocusParameterMonitor FPM
(282.4)
Revision 06 EN 02/2023