Mounting
6.2.4

Mount the device

NOTICE
Damaging/destroying the device
Screws which are too long can damage the screw-in thread.
 When choosing the mounting screws, please ensure that the maximum extension into the device
does not exceed 12 mm.
Mount the device securely in order to prevent a relative movement to the beam axis, reducing a danger
posed by scattered radiation.
Mounting space: Please note that the shutter of the FPM is opened and closed during operation. A com-
plete opening of the shutter has to be ensured. Please make sure that there is enough space to prevent
a collision of the shutter with other parts of the machine as well as to exclude a hazard for the operating
personnel by crushing.
In the base plate of the housing (A) as well as in the side wall (B) there are four fastening threads M6 for
the fixture on a customer specific mounting (see Fig. 6.2 and Fig. 6.3 ). Please mount the housing with four
screws M6. We recommend screws of the strength class 8.8 and a tightening torque of 35 N∙m.
There are also two dowel pin holes Ø6 H7x10 each in the base and in the side panel for precise and reproducible
alignment of the unit.
Fig. 6.1:
Mounting options on the FPM
Revision 06 EN 02/2023