The Command Line „Command - Primes FocusParameterMonitor FPM 60 Original Instructions Manual

Table of Contents

Advertisement

11.2.4
The command line „Command"
The FPM is controlled via the command line in Communication > Free Communication.
The following commands are used during the installation:
profiMeas
svSetup
profiMeas
The profiMeas command sets the measurement mode when connected to the LDS.
The measurement with the PLC (if available) is not affected.
Syntax:
profiMeas 0
normal Measurement
profiMeas 1 xx simulated PLC measurement with setup 1 to 4 (xx: 01 - 04).
The measurement is performed as if the measurement was triggered by the PLC.
profiMeas 1 xx Aa Tt BF cccc pp Mm FTPf
simulated SPS measurement with the parameters defined above
xx: Setup-Number (00 - 04)
A: Evaluation algorithm (a=0 for 2. Moment, a=1 for 86%)
T: Trigger-Mode (0= untriggered, 1= classic, 2=SZ-triggering=single shot)
BF: Beamfind (cccc= Beamfind-Level, pp= Beamfind percentage %)
M: Measuremet mode
0= raw data, background, diffusion compensation measurement
1= raw data
2= underground
3= raw data and immediately afterwards background (without further trigger)
profiMeas 2 z
does not start a measurement, but loads the last measured value from the memory.
Can be used to review the last measurement. (z=0 mess data, z=1 raw data)
svSetup
The svSetup command is used to save and change setups in the FPM. All settings such as window size,
resolution, window position are taken from the LDS. Make sure these are set correctly!
Syntax:
svSetup xx Aa Tt BF cccc pp Mm FTPf
xx: Setup Number (00 - 04)
simulated SPS measurement with the parameters defined above
xx: Setup-Number (00 - 04)
A: Evaluation algorithm (a=0 for 2. Moment, a=1 for 86%)
T: Trigger-Mode (0= untriggered, 1= classic, 2=SZ-triggering=single shot)
BF: Beamfind (cccc= Beamfind-Level, pp= Beamfind percentage %)
M: Measuremet mode
0= raw data, background, diffusion compensation measurement
1= raw data
2= underground
3= raw data and immediately afterwards background (without further trigger)
50
FocusParameterMonitor FPM
Revision 06 EN 02/2023