Appendix | ASCII Communication with Controller
A 3.3.6
Interfaces
A 3.3.6.1
Ethernet IP Settings
IPCONFIG DHCP | (STATIC [<IPAddress> [<Netmask> [<Gateway>]]])
Set Ethernet interface.
- DHCP: IP address and gateway are automatically requested by DHCP . System looks for a LinkLocal address after
approx.. 2 minutes if no DHCP server is available.
- STATIC: Set IP address, net mask and gateway in format xxx.xxx.xxx.xxx.
Values stay the same if no IP address, net mask, and/or gateway is typed in.
A 3.3.6.2
Setting for Ethernet Transmission of Measured Values
MEASTRANSFER [NONE | SERVER/TCP [<PORT>]|(CLIENT/TCP | CLIENT/UDP [<IP> [<Port>]])]
The controller can be operated as a server as well as a client for measurement output via Ethernet.
- NONE: No measurement transmission via Ethernet.
- SERVER/TCP: Controller provides a server for the typed in port, under which the mea-sured values can be sent. This
is only possible via TCP/IP .
- CLIENT/TCP: Controller sends measured values via TCP/IP connection oriented to server. The specifying of the IP
address and server port are required, see
- CLIENT/UDP: Controller sends measured values via UDP/IP connectionless to server.
Specifying the server's IP address and port is required.
- IP: IP address of the server, to which measured values are sent when in client-mode
- Port: Port, to which the server gets assigned to in server-mode or to which the mea-sured values are sent in cli-
ent-mode (min: 1024, max: 65535).
A 3.3.6.3
Setting the RS422 Baud Rate
BAUDRATE [<Baudrate>]
Baud rates can be set in Bps for the RS422 interface:
9600, 115200, 230400, 460800, 691200, 921600, 2000000, 3000000, 4000000
A 3.3.6.4
Change Ethernet / EtherCAT
ETHERMODE [ETHERNET | ETHERCAT]
Select whether the controller starts with Ethernet or EtherCAT mode.
The setting is active after save and reboot the controller only.
A 3.3.6.5
Measurements per frame
MEASCNT_ETH [0 | <count>]
Set the maximum frame number per packet for transmission of measurements via Ethernet.
0: Automatic assignment of the frame number per packet
count: Maximum number of frames per packet (0 ... 350)
A 3.3.6.6
TCP On/Off
TCPKEEPALIVE [ON|OFF]
The parameter can have the following states:
- ON: Enables the function "tcp keep alive"
- OFF: Disables the use of "cp keep alive"
IMS 5x00