Appendix | ASCII Communication with Controller
A 3.3.11 Data Output
A 3.3.11.1 Selection of Digital Output
OUTPUT [NONE | ([RS422] [ETHERNET][ANALOG][ERROROUT])]
- NONE: No measurement value output
- RS422: Output of measurement values via RS422
- ETHERNET: Output of measurement values via Ethernet
- ANALOG: Output of measurement values via analog output
- ERROROUT: Error or status information about the switching outputs
Command starts the measurement value output. If not yet done, connect the controller now to the measurement server.
A 3.3.11.2 Data Output Rate
OUTREDUCEDEVICE [NONE | ([RS422][ANALOG][ETHERNET])]
The number of measured values is reduced using the selected interfaces.
- NONE: No reduction of measurement value output
- RS422: Reduction of measurement value output via RS422
- ANALOG: Reduction of measured values output via analog
- ETHERNET: Reduction of measurement value output via Ethernet
A 3.3.11.3 Reduction Counter of Measurement Value Output
OUTREDUCECOUNT [<Number>]
Reduction counter of measurement value output.
Only every n-th measured value is output. n-1 measured values are canceled.
- Number: 1 ... 3000000 (1 means all frames)
A 3.3.11.4 Error Processing
OUTHOLD [NONE | INFINITE | <Number>]
If no valid measured value can be determined, an error is output.
- NONE: No holding the last measurement value, output of error value
- INFINITE: Infinite holding of the last measurement value
- Number: Holding the last measurement value on the number of measuring cycles; then an error value (maximum of
1024) is output.
IMS 5x00
Page 101