Input - Teledyne T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for T3PM1100:
Table of Contents

Advertisement

Find Quality Products Online at:

Input

Item
Specifications
Voltage
Input type
Current
Voltage
Current
Direct
Measure range
input
Sensor
input
Voltage
Current
Direct input range 5 mA
~ 200 mA
Direct input range 0.5A
Input impedance
~ 20 A
Sensor input
Input range 2.5 V ~ 10 V
(EX1)
Input range 50 mV ~ 2 V
(EX2)
Voltage
Current
Direct input range 5 mA
~ 200 mA
Continuous maximum
allowable input
Direct input range 0.5A
~ 20 A
Sensor
input
Input bandwidth
DC, 0.1 Hz ~ 100kHz
Continuous maximum
600 Vrms, CAT Ⅱ
Common-mode voltage
Line filter
select OFF or ON (cut off frequency of 500 Hz)
select OFF or ON (cut off frequency of 500 Hz)
Frequency filter
Simultaneous conversion voltage and current inputs
Resolution
A/D converter
Maximum conversion rate
200
GlobalTestSupply
www.
T3PM1100 User Manual
Floating input through resistive voltage
divider
Floating input through shunt
15 V, 30 V, 60 V,150 V, 300 V, 600 V
5 mA, 10 mA, 20 mA, 50 mA, 100 mA,
200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A
EX1: 2.5 V, 5 V, 10 V
EX2: 50 mV, 100 mV, 200 mV, 500 mV, 1
V, 2 V
Input resistance: approach 2 MΩ
Input resistance:
approach 505 mΩ
Input resistance:
approach 5 mΩ
Input resistance:
approach 100 kΩ
Input resistance:
approach 20 kΩ
peak value of 1.5 kV or RMS value of 1
kV, whichever is less
peak value of 30 A or
RMS value of 20 A,
whichever is less
peak value of 100 A or
RMS value of 30 A,
whichever is less
peak value less than or equal to 5 times
of the rated range
16bits
Approx. 300kHz
.com