Aoutput[:Normal]:Preset; Aoutput[:Normal]:Rate<X - Teledyne T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for T3PM1100:
Table of Contents

Advertisement

Find Quality Products Online at:
Example
:AOUTPUT:NORMAL:MODE1 FIXED
:AOUTPUT:NORMAL:MODE1?
->:AOUTPUT:NORMAL:MODE1 FIXED
• FIXed = Fixed range mode (default value)
Note
Outputs +5 V when the rated value of each
measurement function is received.
• MANual = Manual range mode
The displayed values of the measurement function
when +5 V and –5 V are output as D/A output can
be set to any values of your choice. This enables
the D/A output to be expanded or reduced for each
channel (D/A zoom).
• COMPare = Comparator mode
By comparing with the comparator limits, this
instrument outputs +5 V, 0 V, or –5 V.

:AOUTput[:NORMal]:PRESet

Sets the D/A output items to their default values.
Description
Syntax
:AOUTput[:NORMal]:PRESet {NORMal|INTEGrate}
Parameter
NORMal
INTEGrate
Example
:AOUTPUT:NORAML:PRESET NORMAL
:AOUTput[:NORMal]:RATE<x>
Sets or returns the maximum and minimum
Description
values for when the D/A output is in manual
range mode. (Remote Control Only)
Syntax
:AOUTput[:NORMal]:RATE<x> {<NRf>,<NRf>}
Query Syntax
:AOUTput[:NORMal]:RATE<x>?
Parameter/
<x>
Return parameter
<NRf>
130