:HARMonics:ORDer

Sets or returns the maximum and minimum
Description
harmonic orders that are analyzed.
Syntax
:HARMonics:ORDer {<NRf>,<NRf>}
Query Syntax
:HARMonics:ORDer?
Parameter/
1st <NRf>
Return parameter
2nd <NRf>
Example
:HARMONICS:ORDER 1,20
:HARMONICS:ORDER?
->:HARMONICS:ORDER 1,20

:HARMonics:THD

Sets or returns the equation used to compute the
Description
THD (total harmonic distortion).
Syntax
:HARMonics:THD {TOTal|FUNDamental}
Query Syntax
:HARMonics:THD?
Parameter/
TOTal
FUNDamental (IEC)
Return parameter
Example
:HARMONICS:THD FUNDAMENTAL
:HARMONICS:THD?
->:HARMONICS:THD FUNDAMENTAL